( g 2.23.14)
If using the Quickie iQ, select Monitor Parameter 10.
followed by selecting Brakes. ( g 2.23.15)
Operate the wheelchair by moving the joystick slight forward. Observe the Brake Current and Brake Voltage. 11.
Refer to the following notes:
Move the joystick further forward, followed by moving it to the left, then right, then reverse. Refer to the fol-12.
The Rhythm and Groove use 12 VDC park brakes. Typical Brake Voltage is 11 VDC.
The Brake Current should be approximately 0.5 amps. Typical values are between 0.4 amps and
0.6 amps.
The Brake Voltage and Brake Current should remain constant.
An increase in Brake Current or Brake Voltage indicates that the park brake is shorted to the motor
windings.
lowing notes: