PULSE 6 SERVICE MANUAL
PAGE 3.1
SEPT 2011SUNRISE MEDICAL
Check that the batteries are fully charged and
in good condition; and check all cables and
connections. Check the connections to the left
motor, look for a loose or damaged connector.
Use the meter to check the resistance across the
two bottom contacts (thicker wires) on the 4-pin
motor connector as shown in (figure 3.1.1). If the
meter reads between 0 to 1.5 ohms, then replace
the controller. If none of the above corrects the
problem, replace the left motor.
Two Bars - Left Motor Disconnected
One Bar - Low Battery Voltage
This code could indicate discharged batteries,
failed batteries, or poor battery connections.
Begin by recharging the batteries and then refer
to Section 1 to check batteries and connections.
Fig 3.1.3
Fig 3.1.1
Otherwise, check the brushes on the left motor
Using a Flat head screwdriver (figure 3.1.2)
remove the brush cap to access the brush (figure
3.1.3) Ensure that they are not excessively worn,
see curvature on brush (figure 3.1.4) Replace as
required.
VR2 Controller Diagnostic Codes
Section 3
Fig 3.1.1
Fig 3.1.2
Fig 3.1.4